5. Press CHA+” button again to change to work time adjustment screen.
Flashing WORK segment indicates that work period can be adjusted.
Use CHB +” or CHB - button to adjust the time in range from 1 � 99
seconds (1 sec. increments).
6. Press CHA+” button again to change to rest time adjustment screen.
Flashing REST segment indicates that rest period can be adjusted. Use
CHB +” or CHB - button to adjust the time in range from 1 � 99 sec�
onds (1 sec. increments).
7. Press CHA+” button again to change to ramp up time adjustment
screen. Flashing RMP segment indicates that ramp up time can be ad�
justed can be adjusted. Use CHB +” or CHB - button to adjust the time
in range 0.3 � 9.9 seconds ( 0.1 sec increments)
8. Press CHA+” button again to change to ramp up time adjustment
screen. Flashing RMP segment indicates that ramp up time can be ad�
justed can be adjusted. Use CHB +” or CHB - button to adjust the time
in range 0.3 � 9.9 seconds ( 0.1 sec increments)
9. Press CHA+” button again to change to output SYN/ALT selection screen
(synchronous or alternating) . Flashing SYN (or ALT) symbol indicates
that output type can be set to synchronous (SYN) or alternating (ALT).
Use CHB +” or CHB - button to select SYN or ALT.
Synchronous (SYN) means that both Channel A and Channel B will give
output at the same time of each Work period.
Alternating (ALT) means that Channel A will work while Channel B will
be o for one Work period and then for the next Work period Channel B
will be on while but Channel A will be o and so on until the end of the
programme time.
10. If SYN (Synchronous) has been selected, press the CHA+” button
again to change to delay time adjustment screen. Flashing DLY symbol
indicates that delay time of channel B can be adjusted. Use CHB +” or
CHB - button to set required time in range 0 � 4 seconds ( 0.1 sec incre
ments).
11. Once delay ( or output synchronization) has been adjusted press SET/ESC
button to save settings and return to Home screen.
If CHA+” or CHB
�” button was pressed again rather than SET/ESC then
the device would loop back to the Hz adjustment screen .
12. Repeat above procedure to readjust custom programme.
Vista de pagina 13