NeuroTrac™ IFC Rehab Operation Manual
34
Table 206: Recommended separation distances between portable and mobile
RF communications equipment and NeuroTrac™ product
The NeuroTrac™ product is intended for use in an electromagnetic environment
in which radiated RF disturbances are controlled. The customer or the user of
the NeuroTrac™ product can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communica�
tions equipment (transmitters) and the NeuroTrac™ product as recommended
below, according to the maximum output power of the communications equip�
ment
Rated maxi-
mum output
power of
transmitter
W
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d =1.2 √P
80 MHz to 800 MHz
d =√1.2 P
800 MHz to 2,5 GHz
d = √2.3 P
0,01 0.12 0.12 0.23
0,1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
aected by absorption and reection from structures, objects and people.