NeuroTrac™ IFC Rehab Operation Manual
35
9. Technical Specication
General
Number of channels: 2 for TENS/NMS and 2 for IFC
Channel isolation:
Depends upon use of 4 independent
transformers.
Output Regulation: Constant Current
Waveform: IFCSymmetrical square biphasic
with zero net DC.
TENS/NMSAsymmetrcal, square, biphasic with
zero net DC.
Energy source: 4 x 1.5V AA non�rechargeable or
4 x 1.2V AA rechargeable batteries.
Response to the
voltage of the batteries: If the voltage goes below 4.4 volts +/� 0.2 the
empty battery symbol will ash on and o once
every second.
If the voltage is below 4.0 volts +/� 0.2
the unit will not turn on.
Open Electrode detect: If an open circuit is detected at the output of the
Channel 1 and 2 or Channel A and B the output
current will be reduced to 0 (for both CH1 and
CH2 or CHA and CHB channel).
Size: L:152mm, W: 89mm, D:35mm
Weight: 0.3 Kg ( including batteries)
Environmental
Specication: Storage and transport
Temp: �10 to +50 degrees Centigrade.
Humidity � 0�90%
Operating
Temp: +10 to +30 degrees Centigrade.
Humidity � 0�90%