NeuroTrac™ IFC Rehab Operation Manual
28
6. Skin Electrodes( Self� adhesive electrodes)
Electrode placement
The placement of electrodes is one of the most important parameters in achiev�
ing eective pain relief using TENS. This is best left to your Physiotherapist or
Physician to advise as to which location is most appropriate.
Quadripolar orientation of electordes for IFC stimulation.
True interferential stimulation requires the use of four electrodes placed on the
body as described in the diagram below. Each pair of electrodes creates a current
ow between them. When two pairs of electrodes are used, the current ows
intersect within the tissue. By making the frequencies for the two channels of
stimulation slightly dierent, an interference or beat” frequency is established
at the point of intersection. This “quadrapolar interferential eect provides a
greater amount of stimulation deeper within the tissue than non�interferential
modes of electrotherapy.
A conceptual drawing of quadripolar ( 4 electrodes) stimulation with
interference taking place within tissue