NeuroTrac™ IFC Rehab Operation Manual
33
Table 204: Guidance and manufacturer’s declaration
– electromagnetic immunity
The NeuroTrac™ product is intended for use in the electromagnetic environment
specied below. The customer or the user of the NeuroTrac™ product should
assure that it is used in such an environment.
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic environment
guidance
Conducted RF
IEC 61000�4�6
Radiated RF
IEC 61000�4�3
3 Vrms
150 kHz to
80 MHz
3 V/m
80 MHz to
2,5 GHz
3 Vrms
150 kHz to 80
MHz
3 V/m
80 MHz to 2,5
GHz
Portable and mobile RF communications
equipment should be used no closer to
any part of the
NeuroTrac™product,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance
d = 1.2 √P 150 kHz to 80 MHz,
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2.5GHz
where P is the maximum output power
rating of the transmitter in watts (W) ac�
cording to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Field strengths from xed RF transmit�
ters, as determined by an electromag�
netic site survey
a
,
should be less than the compliance level
in each frequency range
b
. Interference
may occur in the vicinity of equipment
marked with the
following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
aected by absorption and reection from structures, objects and people.
a
Field strengths from xed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broad�
cast cannot be predicted theoretically with accuracy. To assess the electromagnetic environ�
ment due to xed RF transmitters, an electromagnetic site survey should be considered. If
the measured eld strength in the location in which NeuroTrac™ product is used exceeds
the applicable RF compliance level above, the NeuroTrac™ product should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the NeuroTrac™ product.
b
Over the frequency range 150 kHz to 80 MHz, eld strengths should be less than 3 V/m.