NeuroTrac™ IFC Rehab Operation Manual
16
Description of IFC modes:
1. Abrupt (ABT)
Output with low frequency for a set dwell time and then abrupt change to
high frequency for the same dwell time. Cycle is repeated over programme
duration
2. Ramp (SWP)
Stimulation frequency ramps smoothly from low frequency to high fre
quency for a dened modulation time , then back . Cycle is repeated over
programme duration
3. Intermittent stimulation (WORK/REST)
Intermittent stimulation where output amplitude ramps up in 0.5 sec. to
adjusted level and stimulates for a dened WORK time , then it ramps down
in 0.5 sec. to 0 mA, and stays with no stimulation for a dened REST time.
Cycle is repeated over programme duration.
4. Continuous (CON)
Train of constant value pulses on both channels delivered with no inter�
ruption for entire treatment period
5. Dual Treatment ( ABT / SWP, WORK/REST / CON)
In Programme P06, P07, P10, P11 two modes are combined together into
one programme. Each mode lasts for ½ of the dened total programme
time. When changing from one mode to another the amplitude is reduced
50 % and needs to be readjusted by patient to desired level.